The short version
Start with a single-frequency reading when you need an answer at one operating frequency. Use a sweep when you need the whole story across a band. Then make one physical adjustment, repeat the same measurement, and keep the change only if the curve moves in the direction you wanted.
You can Download the complete CAA 500 Mark II PDF guide for the full button-by-button reference, diagrams, examples, and troubleshooting notes.
Before you measure
Disconnect the radio or any transmitter before connecting the analyzer. Never apply transmitter RF to the CAA 500 Mark II. Ground or discharge static from the coax connector before attaching it, and use the correct connector: SO 239 from 1.8 to 300 MHz, or Type N from 300 to 500 MHz.
A 50 ohm dummy load is a useful first check. In Normal mode, a proper 50 ohm load should show SWR of 1.1 or less. If it does not, pause before blaming the antenna.
Normal mode: SWR and impedance at one frequency
Turn the analyzer on, select the general range with BAND, then dial in the target frequency with FREQ. The display and cross-needle meter show SWR and impedance. Below 190 MHz, the LCD also shows resistance, R, and reactance, X.
- SWR: how closely the system is matched. Lower is generally better.
- Impedance: the total opposition to RF current.
- R: the resistive component, which is often near 50 ohms in a 50 ohm system.
- X: stored energy from inductive or capacitive behavior. The analyzer shows magnitude, not polarity.
The analyzer measures the system at its connector. A long or lossy feedline can change the impedance presented there and can make the SWR look better than the antenna itself. For feedpoint work, get as close to the antenna as practical; for a station check, measure at the station end on purpose.
Finding resonance
In Normal mode, slowly turn FREQ through the expected operating range while watching SWR. The frequency at the lowest observed SWR is the resonant frequency for the measured system. A dip below your target commonly suggests an element that is electrically long; a dip above your target commonly suggests an element that is short. Make small changes and measure again.

Automatic sweep: the fast band picture
From Normal mode, select the amateur-band segment with BAND and press SWEEP CENTER. The analyzer runs a preset sweep, usually in about 25 seconds, and draws the SWR curve for that segment. Look for the dip, its width, and whether your actual operating frequencies land in the useful low-SWR region. Press GRAPH ON OFF to return to Normal mode.
Manual sweep: the zoom lens
- From Normal mode, press GRAPH ON OFF to enter Manual Plot mode.
- Use BAND and FREQ to set a center frequency.
- Press SWEEP CENTER. The center is held and the sweep limits blink.
- Set the desired span with the knobs, then press A P OFF BANDWIDTH to confirm it.
- Turn FREQ slowly through the range to draw the trace. The stated minimum resolution is 1 kHz with gradual rotation.
Manual sweep is especially useful for a narrow operating range, a simplex frequency, or any nonstandard frequency plan. Start wide enough to understand the shape, then narrow the span if you need more detail.
Five-color overlays: compare one change at a time
During a manual sweep, press A P OFF BANDWIDTH to cycle the trace color through red, blue, green, purple, and orange. This lets you keep up to five sweeps on screen while you compare a small change in element length, antenna height, surroundings, radial layout, or matching network position.
Keep the same center and span for a fair comparison. If the dip moves toward the target frequency, that is evidence that the adjustment helped. Photograph or note the trace before leaving graph mode, because graph data is not preserved when you change modes.
Power and battery habits
The analyzer runs from six AA alkaline or NiMH cells, or from an external 8 to 16 volt DC source. With NiMH cells installed, the internal charge switch may be set to ON for slow trickle charging. With alkaline cells, the charge switch must remain OFF. The Auto Power Off timer can be set from 0 to 9 minutes by holding A P OFF in Normal mode and turning BAND; 0 disables the timer.
A simple tuning loop
- Check the analyzer against a dummy load if the reading seems suspect.
- Measure the target frequency in Normal mode.
- Use an automatic or manual sweep to see the larger pattern.
- Make one small, reversible physical change.
- Overlay or repeat the same sweep, then verify the frequencies you actually plan to use.